TracheoSense: Innovation in Endotracheal Tube Cuff Pressure Monitoring with a Three-Dimensional Printed Electronic Manometer-Syringe Device
DOI:
https://doi.org/10.32412/pjohns.v41i1.2383Keywords:
airway management, intubation, printing, three-dimensional, manometry, syringesAbstract
Objective:To design and validate TracheoSense, a 3D-printed manometer-syringe device for measuring endotracheal tube (ETT) cuff pressure, by assessing its accuracy and precision compared with a commercially available standard manometer.
Methods:
Design: Instrument Innovation
Setting: Tertiary Private Training Hospital
Participants: None
Results:The study compared the TracheoSense device with the VBM Analog Manometer, the reference standard, for measuring cuff air volume at target pressures of 20 and 30 cmH2O across tracheal model diameters of 19 mm, 21 mm, and 23 mm. At 20 cmH₂O, mean cuff volumes measured by TracheoSense ranged from 4.04 ± 0.09 mL to 5.82 ± 0.07 mL, with coefficients of variation (CV) between 1.15% and 2.18%, and minimal bias compared to the reference device (−0.02 to 0.09 mL). At 30 cmH₂O, mean volumes ranged from 4.51 ± 0.11 mL to 6.31 ± 0.23 mL, with CVs of 1.62% to 3.58% and biases between −0.07 and 0.02 mL.
Conclusion:The TracheoSense 3D-printed manometer-syringe device demonstrated comparable accuracy and precision in measuring ETT cuff pressures when compared with a commercially available analog manometer. These results support this device as a reliable and practical tool for cuff pressure monitoring.
Downloads
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Publisher

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Copyright transfer (all authors; where the work is not protected by a copyright act e.g. US federal employment at the time of manuscript preparation, and there is no copyright of which ownership can be transferred, a separate statement is hereby submitted by each concerned author). In consideration of the action taken by the Philippine Journal of Otolaryngology Head and Neck Surgery in reviewing and editing this manuscript, I hereby assign, transfer and convey all rights, title and interest in the work, including copyright ownership, to the Philippine Society of Otolaryngology Head and Neck Surgery, Inc. (PSOHNS) in the event that this work is published by the PSOHNS. In making this assignment of ownership, I understand that all accepted manuscripts become the permanent property of the PSOHNS and may not be published elsewhere without written permission from the PSOHNS unless shared under the terms of a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) license.